When it comes to the maintenance of water machines, most water machine operators believe that it is the responsibility of the maintenance department or engineers. In fact, this is not the case. In addition to professional maintenance, it is also necessary to take maintenance measures in the following daily details:
1. Regularly clean the salt box to keep the interior clean
2. Replace the filter element in a timely manner, and clean the precision filter with pure water to prevent the deposition of internal impurities when replacing the filter element.
The replacement time for the filter element is 1-3 months, which can be adjusted according to the actual situation, but the longest cannot exceed 3 months. If the original filter element is relatively clean during replacement, the original replacement cycle can be continued. If the filter element is heavily contaminated, the replacement cycle needs to be shortened.
3. Regular disinfection
The disinfection frequency of the disinfection machine should not exceed once a month, and if the bacterial monitoring results are normal, this disinfection frequency can be executed. If regular disinfection is carried out every month and the bacterial and endotoxin monitoring results exceed the standard, the disinfection frequency needs to be adjusted to every three weeks, two weeks, or one week, and the appropriate disinfection cycle should be determined based on long-term bacterial and endotoxin monitoring results.
The 80 ℃ hot flushing water machine performs weekly hot flushing and chemical disinfection every three months. On this basis, monitor the system for bacteria and endotoxins. If qualified, perform this disinfection cycle. If unqualified, adjust the disinfection frequency, which can be set to hot rinse every three or two days or even every day, and chemical disinfection every month. The level of bacterial growth is different in summer and winter. The temperature in summer is hot, and bacteria grow rapidly. Therefore, disinfection should be strengthened.
The 121 ℃ sterilization water machine can meet the requirements by performing weekly automatic sterilization.
After the implementation of YY0572-2015 blood dialysis and related treatment water standards, increasing disinfection frequency and upgrading water machine disinfection methods have become effective methods to maintain bacteria and endotoxins within the qualified range.
4. Regularly wipe the equipment to keep the surface dry and clean
Regularly wiping the equipment can prevent the adhesion of dust, water, and insect corpses on the surface of the water machine. Generally, it is best to wipe it once every 3-7 days. After wiping, you can apply another layer of metal protective oil to keep the surface of the water machine shiny and new.
5. Daily calibration time for filters (softening, filtering, activated carbon)
6. Test hardness and total chlorine daily, record various indicators of water machine operation (conductivity, water production, pump pressure, etc.)
7. Add salt daily, depending on the dialysis situation
8. Conduct bacterial culture once a month
9. Conduct endotoxin testing once every quarter
10. Conduct an annual chemical pollutant content test